Warning: Whether the screening device or evidentiary breath testing instrument is reliable or unreliable, it is still a criminal offence in Canada and some other jurisdictions to refuse a proper demand by a police officer to provide a breath sample. 

The video below has been suggested by the following DUI DWI Attorney. Click on the Banner for more information about DUI DWI Laws and how an Attorney can help you:

Click here for sponsor

 

Title: Limonene30 - Low & High Cal. Checks Caused by Limonene

Instrument: Intoxilyzer 5000 64 series CA model

Intoxilyzer 5000 in need of maintenance. Experiments with GOJO hand cleaner containing limonene partially dissolved in cold 50 mg/100mls alcohol standard bubbled into breath tube during some air blanks. Effects on cal. checks with 200 alcohol standard. Sometimes but not always "Interferent" flag. Cal. checks far below and far above acceptable. Simulator/tubing/Intoxilyzer reliability compromised because cal. checks no longer reliable. Intoxilyzer® is a registered trademark of CMI, Inc. The Intoxilyzer® 5000C is an "approved instrument" in Canada.

Limonene30

If Youtube embedded videos do not run properly or non-DUI videos appear there may be a problem with your Adobe Flash Player version, click here for Help.

Submitted: 30/04/2009 4:07:34 PM

Last Updated: 20090611015517

 

 

Warning: Whether the screening device or evidentiary breath testing instrument is reliable or unreliable, it is still a criminal offence in Canada and some other jurisdictions to refuse a proper demand by a police officer to provide a breath sample.